Continuously reliable temperature monitoring at extremely low temperatures
The LogTag® CP11S-30 is a precision Smart Probe that can be calibrated separately and easily replaced without moving your data logger. Its plug-and-play hot-swappable design allows for quick replacement, after which the expired probe can be recalibrated at an accredited laboratory. Ideal for applications where continuous, accurate temperature recording is critical, such as laboratories, clinics, and pharmaceutical storage.
Key features
Measuring range: -90 °C to +40 °C (-130 °F to +104 °F)
Exceptional accuracy:
±0.5 °C (-20 °C to +10 °C)
±0.7 °C (-50 °C to -20 °C & +10 °C to +30 °C)
±1 °C (-80 °C to -50 °C & +30 °C to +40 °C)
±1.8 °C (-90 °C to -80 °C)
Temperature resolution: 0.1 °C to 0.5 °C depending on measuring range
Sensor length: 3 m
Hot-swappable & calibratable: replace and calibrate the probe without removing the logger
Connector: high-quality, gold-plated MCX
Protection: IP68 probe tip, IP64 probe/junction, IP44 signal conditioning
Power supply: CR2032 3V LiMnO2 battery (replaceable by distributor)
Battery life: ±3 years of normal use at 5-min logging
Material: Polycarbonate housing, stainless steel probe tip (316)
Probe dimensions: 32.5 x 46.3 x 12.5 mm
Weight: 34 g
Cable material: PTFE
Compatible with: TREL30-16CP loggers
Benefits of the CP11S-30
Continuously reliable temperature recording: ideal for laboratories, clinics, cold storage areas, server rooms, and pharmaceutical or vaccine depots.
Easy replacement: plug in a new probe and continue monitoring without removing the data logger from its location.
Calibration certificate: easy to download via LogTag® Analyzer.
Long-lasting performance: battery with a ±3-year lifespan, minimal maintenance needed.
Applications
Laboratories & clinics
Cold storage areas & freezers
Server rooms & data centers
Pharmaceutical & vaccine storage
Environments where accurate, continuous monitoring is essential